PR 2559: Dewar 3 iced up and occasionally complaining of ln2temp, ln2 empty The blue dewar on spectrograph two ( CCD03 ) has a large ice build-up. About every hour and a half there is an error that the ln2 is empty along with an ln2temp error. Looking at the ln2temp plot one sees five or six data points around the nominal temperature of 92 K, followed by one point somewhere between 110 and 120 K. It looks like the fill interval is just slightly longer than the hold time. None of the other dewars are showing this behavior. Is is possible the vacuum isn't what it should be, or maybe there is a leak in the fill tube where it connects to the dewar, leading to inefficient or incomplete fills? PR 2560: Twice in two days M2 actuator E couldn't stop on a full step. In the last two days, each time early in the evening, we started getting errors from the M2 galil when it attempted to adjust the collimation of the secondary during tracking. The galil was complaining that it could not get actuator E to stop on a full step. 0 0 F Modu="mir_o_MoveOneGalil"; Text=" Sec returned an error message" 0 0 I Received="?GOCORR one or more axes did not end on a full step 0, 0, 0, 0, 1" 0 0 W Text="failed to adjust collimation and focus" In both cases homing the E axis restored operation and it kept working the rest of the night.